Analysis of historical records orthofix (b)(4) checked the internal records related to the controls made on the device code 99-93100 lot b1069969 before the market release.No anomalies have been found.The original lot, manufactured in 2017, was comprised of (b)(4) devices.All of them have already been distributed to the market.According to orthofix (b)(4) historical records, this is the first complaint notified from this specific device lot.Technical evaluation the device involved in this event was not returned to orthofix (b)(4) as it was discarded by the hospital.It was therefore not possible to perform the technical investigation on the involved device.Medical evaluation the information made available on the case was sent to our medical evaluator.Please see below an extract of the medical evaluation: "this patient is a (b)(6) year old male with a subcapital proximal humeral fracture.The patient is of normal height and weight and we can expect the bone to be strong.The fracture was being fixed in reduction with the galaxy threaded wires.One of the wires broke during insertion and the broken part removed and discarded.Therefore we will not be able to test it.The likely cause is excessive drill speed with healthy hard bone." final comments the device involved in this event was not returned to orthofix (b)(4) as it was discarded by the hospital.It was therefore not possible to perform the technical investigation on the involved device.The medical evaluation evidenced that: "this patient is a (b)(6) year old male with a subcapital proximal humeral fracture.The patient is of normal height and weight and we can expect the bone to be strong.The fracture was being fixed in reduction with the galaxy threaded wires.One of the wires broke during insertion and the broken part removed and discarded.Therefore we will not be able to test it.The likely cause is excessive drill speed with healthy hard bone." the orthofix (b)(4) operative technique galaxy fixation system upper extremities, code ref.Gf 1101 opt, on page 25, indicates the procedure to insert the wires: "inserting the wires with the aid of the wire targeting device: insert the wires at slow speed.Position the first wire using the soft tissues protective guide.The correct position of the wires must be confirmed by x-rays." based on the results the event investigation, orthofix (b)(4) can conclude that the problem occurred could be attributable to the conditions of use of the device.It could have happened that the wire was inserted at an excessive drill speed or with a lateral flexion.These application modes can significantly increase the torsional and flexional load, leading to breakage.The analysis of the historical data evidenced that no other similar notifications have been received on devices belonging to the same lot.Orthofix (b)(4) continues monitoring the devices on the market.
|
|
The information provided by the local distributor indicates: hospital name: (b)(6); surgeon name: dr.(b)(6); date of surgery: (b)(6) 2018; body part to which device was applied: left proximal humerus; surgery description: fracture treatment; patient's information: (b)(6) year-old, male, weight (b)(6) kg, height 165 cm, previous health condition: good; problem observed during clinical use on patient; event description: "the wire broke during insertion.The surgeon was able to remove the portion of the wire in the humerus." notes: the broken wire was discarded.The complaint report form also indicates: -the device failure had no adverse effects on patient.-the initial surgery was completed with the device.-the event did not lead to a clinically relevant increase in the duration of the surgical procedure.-an additional surgery was not required.-copy of operative reports is not available.-copy of x-ray images is available.-information on patient current health condition: good.On the same date of notification, orthofix (b)(4) receives copy of post-operative x-ray images.Manufacturer reference number: (b)(4).
|