Catalog Number RONYX27526UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 01/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A resolute onyx drug eluting stent was intended to be used to treat a heavily calcified lesion in the rca.Tortuosity was moderate.Proximal and distal previous stents were present.The distal stent had in-stent restenosis (isr).Negative prep was not performed.The lesion was not pre-dilated.An attempt was made to directly stent the lesion with the resolute onyx.The device was passed through the previously-deployed stent.Resistance was encountered when advancing the device.The device could not cross the distal in-stent lesion.The attempt was a little forceful, but excessive force was not used.The stent catheter was removed and it was noted that the stent had dislodged and was in the vessel.The stent could not be snared.Another stent was used to intentionally trap the resolute onyx stent against the previous stent.The patient was unharmed.Patient status is alive with no injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7356717
Report Number9612164-2018-00531
Device Sequence Number19338838
Product Code NIQ
UDI-Device Identifier00643169556928
UDI-Public00643169556928
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2019
Device Catalogue NumberRONYX27526UX
Device Lot Number0008784744
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/21/2018
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date03/21/2018
Supplement Report FDA Received Date06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-