Model Number 70303QC20
Medical Device Problem Code Sticking (1597)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 02/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that after compacting a train of extension, the seed train hung in the end of the quicklink.It was noted that the tip of the extension was crushed.Two attempts later, the expansion bar was changed to eliminate a batch problem, however the issue persisted.The quicklink was then changed, and so was the extension bar, in addition to compressing the grain train at a slower pace.
 
Event or Problem Description
It was reported that after compacting a train of extension, the seed train hung in the end of the quicklink.It was noted that the tip of the extension was crushed.Two attempts later, the expansion bar was changed to eliminate a batch problem, however the issue persisted.The quicklink was then changed, and so was the extension bar, in addition to compressing the grain train at a slower pace.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿in the event the quicklink¿ loader or cartridges become inoperable due to damage or malfunction, any or all components may be removed from the cartridges and implanted manually." the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKLINK¿ CARTRIDGE 5.0MM EXTENSION SOURCELINK¿
Common Device Name
BRACHYTHERAPY
Manufacturer (Section D)
BARD BRACHYTHERAPY, INC. -1424526
295 east lies road
carol stream IL 60188
MDR Report Key7356729
Report Number1018233-2018-00893
Device Sequence Number19113675
Product Code KXK
UDI-Device Identifier10801741078115
UDI-Public(01)10801741078115
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K023210
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2018
Device Model Number70303QC20
Device Catalogue Number70303QC20
Device Lot NumberBBBV0102
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2018
Supplement Date Received by Manufacturer05/18/2018
Initial Report FDA Received Date03/21/2018
Supplement Report FDA Received Date05/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
Patient Weight87
-
-