• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a call service alarm (cs 069) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Additional Manufacturer Narrative
Follow-up # 1 date of submission (b)(6) 2018 device evaluation: the device has been returned and evaluated by product analysis on (b)(6) 2018 with the following findings: during testing, the pump was powered on and emitted a call service 069 call service alarm immediately.The call service alarm was recorded in the black box data as a cs 087 failure, indicating a language file corruption due to a failed u39 component on the printed circuit board.Unrelated to the initial complaint, it was observed that the battery compartment had two cracks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7357450
Report Number2531779-2018-05781
Device Sequence Number1220858
Product Code OYC
UDI-Device Identifier10840406100099
UDI-Public10840406100099
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Was the Report Sent to FDA? Yes
Device Age27 MO
Initial Date Received by Manufacturer 03/01/2018
Supplement Date Received by Manufacturer03/01/2018
Initial Report FDA Received Date03/21/2018
Supplement Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age18 YR
-
-