| Model Number |
TV-AB3480-E |
| Medical Device Problem Codes |
Failure To Adhere Or Bond (1031); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Endologix will continue to investigate the reported event.The patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.A final report will be submitted once the investigation of the reported event has concluded.
|
| |
|
Event or Problem Description
|
|
An ovation prime abdominal stent graft system was implanted to treat an abdominal aortic aneurysm.The patient returned for an unknown reason where physician observed a type iiia endoleak due to iliac limb separation.The physician has elected to re-intervene, however, a date for the re-intervention has yet to be scheduled.As of the date of this report, there have been no additional patient sequelae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the clinical assessment for this event, the most likely cause of the loss of seal was found to be unknown/undetermined due to the lack of medical records and imaging surrounding the event.Procedure related harms and the final patient disposition could not be determined as well, however, there have been no additional reports of patient sequelae.A review of the device quality records shows that the device demonstrated compliance to established procedures and specifications at the time of manufacture.
|
| |
|