| Catalog Number |
5517F401 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate that devices were manufactured and accepted into final stock on with no relevant reported discrepancies.There have been no other events for the reported lot.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
When the implant was opened.It was noticed that the internal packaging damaged.
|
| |
|
Event or Problem Description
|
|
When the implant was opened.It was noticed that the internal packaging damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding packaging damage involving a triathlon femoral component was reported.The event was confirmed by visual inspection.Conclusion: based on the visual inspection of the returned product the investigation concluded that the packaging damage was most likely caused by incorrect/excessive handling prior to opening of the product whereby the unit carton may have been compressed and/or dropped from a height causing the damage to the carton.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|