Model Number AC0201250
Health Effect - Clinical Code Infiltration into Tissue (1931)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of recn0300 showed two other similar product complaint(s) from this lot number.All three complaints for this lot number recn0300 have been reported from the same facility.
 
Event or Problem Description
It was reported that: "catheter inserted in ac space of the arm, flushed easily and had reliable blood return, patient received a ct scan with power contrast at 300 psi and 3.2 ml/sec and part of the injection was successful and but a significant amount of the contrast was injected outside of the vein and caused an infiltration.Test was still successful.We speculated that not much of the catheter was in the vein and the pressure from the contrast injection forced the catheter back and off of the vessel." no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUCATH, INTRAVASCULAR CATHETER 20 GX1.25
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7357768
Report Number3006260740-2018-00474
Device Sequence Number2156143
Product Code FOZ
UDI-Device Identifier00801741110917
UDI-Public(01)00801741110917
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K153298
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAC0201250
Device Catalogue NumberAC0201250
Device Lot NumberRECN0300
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-