| Model Number |
OMS-A8 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a kidney transplant procedure, the device had issues with the loading or firing of the clips, wherein the clips were unable to be loaded properly into the jaws at any point during use.A new device was opened in order to resolve the issue.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with sixteen clips remaining.There were no visual or functional abnormalities noted during pmv testing.The remaining clips were fired on test media with proper formation.The jaw and handle moved smoothly through the firing cycle and returned to the open position and each remaining clip loaded properly in the jaw.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
|
| |
|