It was reported that during an abdomen with runoff procedure, the flexor ansel guiding sheath hub broke off upon removal.There were no devices being utilized through the sheath at the time of the separation.The sheath was placed in the patient's leg via a contralateral approach.A wire guide (manufacturer unspecified) was inserted into the lumen of the sheath prior to removal.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence. according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation ¿ evaluation.A review of the complaint history, device history record, documentation, manufacturing instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the hub was separated from the sheath of the device.The proximal fittings were disassembled to verify if any sheath material was present.A 2 mm compression was present 2.5 cm from the proximal end.The sheath flare appeared to be deformed, and in an out-of-round shape.It was examined with a flare gauge, and it failed the inspection as it did not pass through either lumen on the flare gauge.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation a review of the complaint history, device history record, documentation, manufacturing instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the hub was separated from the sheath of the device.The proximal fittings were disassembled to verify if any sheath material was present.A 2 mm compression was present 2.5 cm from the proximal end.The sheath flare appeared to be deformed, and in an out-of-round shape.It was examined with a flare gauge, and it failed the inspection as it did not pass through either lumen on the flare gauge.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|