• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT0408
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pusher assembly was kinked approximately 142.0 and 143.0 cm from the proximal end.The embolization coil was intact with the pusher assembly and had offset coil winds near its proximal end.Conclusions: evaluation of the returned device revealed the embolization coil had offset coil winds near its proximal end.If the introducer sheath is not properly aligned in the hub prior to advancement, resistance may be experienced, and damage such as this may occur.The pusher assembly kinks likely occurred due to forceful advancement of the smart coil against this resistance.The non-penumbra microcatheter mentioned in the complaint was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the internal carotid-posterior communicating artery using penumbra smart coils (smart coils).During the procedure, the physician successfully placed two coils using a non-penumbra microcatheter.The physician then experienced resistance and was unable to advance a smart coil through the hub of the microcatheter, and therefore the smart coil was removed.The physician re-attempted to advance the same smart coil into the microcatheter, however the same resistance was felt again.The physician therefore removed the smart coil and completed the procedure using additional coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7358417
Report Number3005168196-2018-00624
Device Sequence Number709158
Product Code HCG
UDI-Device Identifier00814548015392
UDI-Public00814548015392
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/20/2022
Device Catalogue Number400SMTXSFT0408
Device Lot NumberF78761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2018
Initial Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-