• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (CE) ANGIOJET® ULTRA SYSTEM CONSOLE; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (CE) ANGIOJET® ULTRA SYSTEM CONSOLE; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 105650-001R
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
 
Event or Problem Description
It was reported that loss of power occurred to the console.An angiojet ultra system console was selected for use.It was reported that during thrombectomy mode, the console lost power.It was not specified which type of catheter was being used at the time of occurrence.No patient complications were reported and the procedure was completed with another same device.
 
Additional Manufacturer Narrative
Update: device evaluated by mfr., eval summary attached, method and conclusion codes.Device evaluated by manufacturer: the ultra system with no catheter was received in fair condition with no physical damages/defects observed.Functional testing was performed and the user interface showed a blank display on startup sequence during testing.The ultra main power at the back of the system and barcode assembly light was on but the main control power will not turn on.The ultra system console was reset twice but the main control power failed.A service history review was performed and nothing was found to indicate a possible service-related cause for the complaint.The investigation conclusion isolated the failure to a specific component within a bsc system or assembly which contributed to the event; however the cause for the component failure is unknown.(b)(4).
 
Event or Problem Description
It was reported that loss of power occurred to the console.An angiojet ultra system console was selected for use.It was reported that during thrombectomy mode, the console lost power.It was not specified which type of catheter was being used at the time of occurrence.No patient complications were reported and the procedure was completed with another same device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET® ULTRA SYSTEM CONSOLE
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
MDR Report Key7359587
Report Number2134265-2018-02892
Device Sequence Number1392809
Product Code MCX
UDI-Device Identifier08714729890607
UDI-Public08714729890607
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P980037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number105650-001R
Device Catalogue Number105650-001R
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2018
Supplement Date Received by Manufacturer03/26/2018
Initial Report FDA Received Date03/21/2018
Supplement Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-