| Model Number |
X SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The malfunction was duplicated and the defib receptacle assembly was replaced to resolve the malfunction.Analysis of reports of this type has not identified an increase in trend.
|
| |
|