• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DHS®/DCS® GUIDE SHAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DHS®/DCS® GUIDE SHAFT Back to Search Results
Catalog Number 338.21
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient information is unknown.Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Without a lot number the device history records review could not be completed.The manufacture date is unknown.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device malfunction did occur and regardless of misuse, has caused or contributed to the need for medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.The complaint indicated that the devices malfunctioned which resulted in a change in surgical plan (planned procedure was dhs, changed to dhhs) and a surgical delay of 8-10 minutes.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a unknown procedure on (b)(6) 2018, while the surgeon was using the dhs®/dcs® wrench, for inserting a dhs®/dcs® coupling screw, upon insertion into the femoral head, the dhs®/dcs® wrench would not disconnect and slide off from guide shaft assembly (dhs®/dcs® centering sleeve and dhs®/dcs® guide shaft).Dhs®/dcs® lag screw 12.7mm thread/95mm (part # 280.950, lot # 5522496, qty # 1) was attached to the guide shaft and coupling screw and was not able to detach because it was stuck inside of the wrench.Upon trying to disconnect it, the handle broke off the dhs wrench.The dhs®/dcs® coupling screw and the guide shaft assembly were removed by moving back and forth.There was a surgical delay of 8-10 minutes.The surgery was completed using the dynamic helical hip system (dhhs).The patient was unharmed.Concomitant devices reported: dhs®/dcs® coupling screw (part # 338.20, lot # unknown, qty #1).This is report 2 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A product investigation was performed.The returned guide shaft was received unable to be disassembled from the lag screw (280.950), centering sleeve (338.190) and wrench (338.06).The instrument can¿t be seen visually as its unable to be disassembled from the rest of the assembly instrument.The complaint condition could be replicated.It is likely that possible rough handling and/or cross threading during surgery has led to this complaint condition.The complaint is confirmed.Since the device can¿t be seen not lot number could be determined and therefore a device history could not be reviewed.And since the device can¿t be seen no dimensional analysis and material properties can be performed.Drawing smw_100626 rev a was reviewed and no design issues or discrepancies were identified.Risk assessment review: the drill/protection guides/sleeves design and clinical risk management (dcrm) document, windchill 0000262102 a.2, was reviewed in action a-741027 and found to adequately address the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DHS®/DCS® GUIDE SHAFT
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7360196
Report Number2939274-2018-51146
Device Sequence Number581081
Product Code FZX
UDI-Device Identifier10886982193247
UDI-Public(01)10886982193247(10)UNKNOWNLOTNUMBER
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number338.21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Initial Date Received by Manufacturer 02/22/2018
Supplement Date Received by Manufacturer04/02/2018
05/14/2018
Initial Report FDA Received Date03/21/2018
Supplement Report FDA Received Date04/19/2018
05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight109
-
-