| Model Number |
2261 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Therapeutic Response, Increased (2272)
|
| Date of Event |
02/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Instrumentation laboratory (il) conducted an investigation that included a review of the data files from the customer's gem premier 4000.The data review showed that the device was operating within specifications.There was no system or sensor malfunction during analysis of the patient's sample.Although the patient's sodium intake was increased (dosage not known), there was no perceived impact to the patient.We are submitting this mdr due to the location (nicu) of the patient.No remedial action is required.
|
| |
|
Event or Problem Description
|
|
Customer reported that due to a sodium value of 128 mmol/l on gem premier 4000 located in the nicu, the patient's sodium intake was increased.However, when the lab analyzer reported a sodium value of 150 ten hours later, the gem premier 4000 sodium result was questioned.There was no perceived impact to the patient due to the increased sodium intake.
|
| |
|