Model Number MICL12.6
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: lens was returned in liquid, in lens vial.Visual inspection found a missing piece of haptic.No similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that a 12.6mm micl12.6 implantable collamer lens, -10.0 diopter would not load properly in the cartridge.No harm to the patient, no patient contact.The back up lens was used with no issues.Cause of event was unknown.
 
Additional Manufacturer Narrative
Corrected data: the reporter indicated that a 12.6mm micl12.6 implantable collamer lens of -10.0 diopter tore during loading as it would not load properly in the cartridge.There was no harm to the patient and no patient contact.The back-up lens was used with no issues.The cause of the event was unknown.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7361609
Report Number2023826-2018-00457
Device Sequence Number13479408
Product Code MTA
UDI-Device Identifier00841542103022
UDI-Public00841542103022
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/30/2019
Device Model NumberMICL12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/21/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-