| Model Number |
MICL12.6 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: lens was returned in liquid, in lens vial.Visual inspection found a missing piece of haptic.No similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm micl12.6 implantable collamer lens, -10.0 diopter would not load properly in the cartridge.No harm to the patient, no patient contact.The back up lens was used with no issues.Cause of event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: the reporter indicated that a 12.6mm micl12.6 implantable collamer lens of -10.0 diopter tore during loading as it would not load properly in the cartridge.There was no harm to the patient and no patient contact.The back-up lens was used with no issues.The cause of the event was unknown.Claim# (b)(4).
|
| |
|