• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM CORTEX SCREW SELF-TAPPING 26MM; SCREW,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM CORTEX SCREW SELF-TAPPING 26MM; SCREW,FIXATION,BONE Back to Search Results
Model Number 214.826
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narrative: (b)(6).Patient weight was not provided for reporting.Date of event is unknown.(b)(4).Date of implant is unknown.Date of event was reported as (b)(6) 2018.It is unknown if this date represents date of revision/explant.Device is not expected to be returned for manufacturer review/investigation.Device evaluation/investigation could not be completed; no conclusion could be drawn as product was not returned to manufacturer.Additionally, device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, patient underwent initial open reduction and internal fixation (orif) humeral shaft surgery and the patient was implanted with a plate and seven (7) screws.Patient underwent revision of orif humeral shaft due to loosening of the proximal screws from the plate causing a nonunion of the fracture.During the revision, the plate and screws were removed, and a humeral nail and three (3) locking screws were implanted to stabilize the fracture.Concomitant device reported: 4.5mm narrow lcp® plate 10 holes/188mm (part# 224.601, lot# unknown, quantity 1).This report is for one (1) 4.5mm cortex screw self-tapping 26mm.(b)(4).
 
Additional Manufacturer Narrative
Patient code (b)(4) used in the initial report to capture additional medical/surgical intervention required.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM CORTEX SCREW SELF-TAPPING 26MM
Common Device Name
SCREW,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7361637
Report Number2939274-2018-51165
Device Sequence Number10795303
Product Code HWC
UDI-Device Identifier10886982154620
UDI-Public(01)10886982154620(10)LOT#UNKNOWN
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K112583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number214.826
Device Catalogue Number214.826
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer03/22/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date03/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-