• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 5 10MM; KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 5 10MM; KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151620510
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Investigation summary: evaluation of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision for: pain on tibial side , after 3 months ok.When tibia base plate inspected intra-operatively found to be loose, when removed from patient found no cement on implant.Poly wear.Doi: (b)(6) 2016; dor: (b)(6) 2018; unknown knee.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: evaluation of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ 5 10MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key7361664
Report Number1818910-2018-55871
Device Sequence Number18875297
Product Code JWH
UDI-Device Identifier10603295045687
UDI-Public10603295045687
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151620510
Device Lot Number537968
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer04/25/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date05/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-