COOK IRELAND LTD ZIMMON BILIARY STENT SET; FGE CATHETER, BILIARY, DIAGNOSTIC - BILIARY STENT - METAL
|
Back to Search Results |
|
| Model Number |
G21393 |
| Medical Device Problem Code |
Kinked (1339)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
A (b)(6) year old lady patient, had a procedure for a fistula on gastrostomy sleeve.At the beginning of the procedure, the stent folded in its packaging and staff could not unfold it.A new stent was used.The customer wants a samson box, or the device will be destroyed.Customer requests a free replacement or a credit note.Lot number to be confirmed by cust, as cust wrote lot c142921 on enclose document.I took the lot that corresponds the most."as per complaint form": the nurse says the stent was kinked in the package; reporting based on kink precedence for this device family.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to mfr site as follows: importer site contact and address: (b)(6).(b)(4).Lab evaluation: 1 x zebd-7-4 of lot#: c1421921 was returned to cirl for evaluation.Upon evaluation of the returned device, a 0.035 inch wire guide could not pass porthole 3.The pigtail straightener was returned with the device.There were no other defects observed.Root cause: a definitive root cause for the customer complaint could not be determined, as the exact operational conditions of use could not be replicated in a laboratory setting.A possible root cause may be that the device may have become damaged during transportation and storage.However, as the exact transport or storage conditions cannot be replicated in a laboratory setting, this cannot be conclusively determined.The complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Documents review: a review of the manufacturing records for the zebd-7-4 device of lot#: c1421921 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot#: c1421921; upon review of complaints, this failure mode has not occurred previously with this lot#: c1421921.Fqc/pkg review: prior to distribution, all zebd devices are subject to visual inspection and functional checks to ensure device integrity.There is 100% inspection on all stents for kinks and a check that the appropriate size wire guide moves smoothly and freely when inserted into both ends of the stent.There is also a visual inspection of the product and packaging at packaging, packaging qc and post sterile qc, "complete a 100% visual inspection of the packaged unit (tyvek pouched) while held at a comfortable arm¿s length from the unaided eye at normal lighting conditions." ifu review: it may be noted that according to the instructions for use the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.Summary: the complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
Report is being submitted under the precedence for "stent kinked".58 years old lady patient, had a procedure for a fistula on gastrostomy sleeve.At the beginning of the procedure, the stent folded in its packaging and staff could not unfold it.A new stent was used."as per complaint form": the nurse says the stent was kinked in the package.
|
| |
|
Event or Problem Description
|
|
A 58 years old lady patient, had a procedure for a fistula on gastrostomy sleeve.At the beginning of the procedure, the stent folded in its packaging and staff could not unfold it.A new stent was used."as per complaint form": the nurse says the stent was kinked in the package; reporting based on kink precedence for this device family.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Importer site contact and address: (b)(4), cook medical incorporated (cmi), 1025 acuff road, p.O box 4195, bloomington , indiana , 47402-4195.Importer site establishment registration number: 3005580113.Problem statement: the customer reported the following complaint issue: ¿a 58 years old lady patient had a procedure for a fistula on gastrostomy sleeve.At the beginning of the procedure, the stent folded in its packaging and staff could not unfold it.A new stent was used."as per complaint form": the nurse says the stent was kinked in the package.¿ the kink in the device was noted prior to use.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.The lot number c1421921 was confirmed on the complaint information form attached to this complaint.Lab evaluation: 1 x zebd-7-4 of lot # c1421921 was involved in this complaint.The device has not yet been returned for evaluation; therefore, a document-based investigation was completed for this complaint.The investigation evaluation will be updated when the device is returned for evaluation.Root cause: a definitive root cause for the customer complaint could not be determined, as the exact operational conditions of use could not be replicated in a laboratory setting.A possible root cause may be that the device may have become damaged during transportation and storage.However, as the device was not returned for evaluation and the exact transport or storage conditions cannot be replicated in a laboratory setting, this cannot be conclusively determined.The complaint is confirmed based on the customer¿s testimony.Documents review: a review of the manufacturing records for the zebd-7-4 device of lot # c1421921 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1421921; upon review of complaints, this failure mode has not occurred previously with this lot # c1421921.Fqc/pkg review: prior to distribution, all zebd devices are subject to visual inspection and functional checks to ensure device integrity.There is 100% inspection on all stents for kinks and a check that the appropriate size wire guide moves smoothly and freely when inserted into both ends of the stent.There is also a visual inspection of the product and packaging at packaging, packaging qc, and post sterile qc as per qsi0975 "complete a 100% visual inspection of the packaged unit (tyvek pouched) while held at a comfortable arm¿s length from the unaided eye at normal lighting conditions." ifu review: it may be noted that according to the instructions for use, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.Summary: the complaint is confirmed based on the customer¿s testimony.Complaints of this nature will continue to be monitored for potential emerging trends.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Importer site contact and address: (b)(4), cook medical incorporated (cmi), 1025 acuff road, p.O box 4195, bloomington , indiana , 47402-4195.Importer site establishment registration number: 3005580113.Problem statement: the customer reported the following complaint issue: ¿a 58 years old lady patient had a procedure for a fistula on gastrostomy sleeve.At the beginning of the procedure, the stent folded in its packaging and staff could not unfold it.A new stent was used."as per complaint form": the nurse says the stent was kinked in the package.¿ the kink in the device was noted prior to use.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.The lot number c1421921 was confirmed on the complaint information form attached to this complaint.Lab evaluation: 1 x zebd-7-4 of lot # c1421921 was involved in this complaint.The device has not yet been returned for evaluation; therefore, a document-based investigation was completed for this complaint.The investigation evaluation will be updated when the device is returned for evaluation.Root cause: a definitive root cause for the customer complaint could not be determined, as the exact operational conditions of use could not be replicated in a laboratory setting.A possible root cause may be that the device may have become damaged during transportation and storage.However, as the device was not returned for evaluation and the exact transport or storage conditions cannot be replicated in a laboratory setting, this cannot be conclusively determined.The complaint is confirmed based on the customer¿s testimony.Documents review: a review of the manufacturing records for the zebd-7-4 device of lot # c1421921 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1421921; upon review of complaints, this failure mode has not occurred previously with this lot # c1421921.Fqc/pkg review: prior to distribution, all zebd devices are subject to visual inspection and functional checks to ensure device integrity.There is 100% inspection on all stents for kinks and a check that the appropriate size wire guide moves smoothly and freely when inserted into both ends of the stent.There is also a visual inspection of the product and packaging at packaging, packaging qc, and post sterile qc as per qsi0975 "complete a 100% visual inspection of the packaged unit (tyvek pouched) while held at a comfortable arm¿s length from the unaided eye at normal lighting conditions." ifu review: it may be noted that according to the instructions for use, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.Summary: the complaint is confirmed based on the customer¿s testimony.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to (b)(4).Importer site establishment registration number: (b)(4).Following device return a follow up will be sent to update the file once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
A 58 years old lady patient, had a procedure for a fistula on gastrostomy sleeve.At the beginning of the procedure, the stent folded in its packaging and staff could not unfold it.A new stent was used."as per complaint form": the nurse says the stent was kinked in the package; reporting based on kink precedence for this device family.
|
| |
|