Catalog Number W40T0406S
Medical Device Problem Codes Occlusion Within Device (1423); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Method: actual device not evaluated - the device was not returned to vascutek as it was not available for return.All remaining devices from the batch have been sold.Manufacturing review - a complete review of the manufacturing and qc records was carried out.No issues were identified photographic inspection - images provided by the complainant were reviewed by vascutek clinical department associate.Conclusion - procedure related event.Results: operational problems.Conclusion: use error caused or contributed to event.Device not returned.A review of manufacturing and qc records confirmed that the device was manufactured to specification.A review of similar events was completed for 5-year period which gave a low occurrence rate of (b)(5).This is first complaint for occlusion failure for taperflo graft.No other complaints have been received for other units of the same batch.No issues or adverse trends were noted during the preliminary investigation.As the explanted graft was not available for evaluation the images of the incident were provided.Images were reviewed by vascutek clinical department and it was determined that judging by the information provided the event is procedure related.The graft was crushed flat possibly due to puncture technique not being followed.On the image of the explanted graft puncture holes are noticeable on both sides.Puncture of both graft walls might have caused restricted/if any blood flow which would probably flatten the graft and could have contributed to the jelly like substance, most probable thrombus formation.As the event was deemed to be procedure related, no corrective actions were assigned.No further action is required.Vascutek consider this case closed.
 
Event or Problem Description
Terumo (b)(4) reported graft occlusion: the 10th day after implant, puncturing was performed since the condition of the patient looked good with no swelling.The patient was then discharged from the hospital and sent back to the clinic in which the patient had been hospitalized originally.The 37th day after implant, the patient was sent back to the hospital again because the graft was occluded.When the graft was explanted, the following were found: jelly-like substances surrounding the graft did not adhere to the graft.The graft was flat, and when the graft was punctured with a needle, the needle perforated the front and rear walls of the graft.A low-echoic area of seroma-like substances was observed around the entire length of the graft which was crushed flatly.When the graft was replaced with a thoratec graft, the implanted graft was easily removed without adhering to the tissue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEALED EPTFE
Common Device Name
TAPERFLO
Manufacturer (Section D)
VASCUTEK LTD
newmains avenue
inchinnan business park
renfrewshire, PA4 9 RR
UK  PA4 9RR
MDR Report Key7361901
Report Number9612515-2018-00002
Device Sequence Number13757089
Product Code DSY
UDI-Device Identifier05037881009599
UDI-Public05037881009599
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K092863
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/22/2018
Report Date (Section F) 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Catalogue NumberW40T0406S
Device Lot Number1590465-9805
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/21/2018
Device Age12 MO
Event Location Hospital
Date Report to Manufacturer02/21/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2018
Initial Report FDA Received Date03/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-