• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE Back to Search Results
Model Number 295251
Medical Device Problem Code Cut In Material (2454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer states an error e6 occurred.Upon triage it was noted that there were exposed copper wires on the power cord.
 
Additional Manufacturer Narrative
An evaluation of the scd 700 was performed for the reported condition of, ¿an error e6 occurred¿.The unit was triaged and the customer¿s reported condition was confirmed and it was found the power cord was damaged.Device cleaned and disinfected according to manufacturer's guideline.Customer's complaint confirmed, the power cord has damaged and was replaced.The device fully checked for functionality and patient electrical safety.The pump was excessively damaged, so the root cause is categorized as customer misuse.A review of the device history record shows that this unit was manufactured and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Common Device Name
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7362558
Report Number3006451981-2018-00207
Device Sequence Number1233702
Product Code JOW
UDI-Device Identifier10884521129580
UDI-Public10884521129580
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number295251
Device Catalogue Number295251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2018
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer03/06/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date03/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-