MAUDE Adverse Event Report: ENCORE MEDICAL L.P. BIOLOX DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM -4.0

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. BIOLOX DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM -4.0 Back to Search Results
Model Number 430-98-052
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/14/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a dislocation.The surgeon only replaced a djo head and the other implants were replaced with a competitor part.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a dislocation.The previous surgery and the revision detailed in this investigation occurred 1.8 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a dislocation.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are many factors that may contribute to the event that are outside the control of djo surgical such as: inadequate soft tissue support, excessive range of motion, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA HIP
Common Device Name
HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM -4.0
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7362724
Report Number1644408-2018-00252
Device Sequence Number13781591
Product Code LZO
UDI-Device Identifier00888912084024
UDI-Public(01)00888912084024
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K082844
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/26/2021
Device Model Number430-98-052
Device Catalogue Number400-03-361
Device Lot Number864B1190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/14/2018
Supplement Date Received by Manufacturer04/03/2018
05/02/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date04/03/2018
05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
010-55-035, LOT 009A1122; 010-55-035, LOT 009A1122; 400-03-361, LOT 864B1190; 430-98-052, LOT 587G1286; 932-36-752, LOT 813N1998; 932-36-752, LOT 813N1998; 010-55-035, LOT 009A1122; 400-03-361, LOT 864B1190; 932-36-752, LOT 813N1998
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-