The insulin pump passed the displacement test.No pump error noted.The insulin pump had compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or test for motor error alarm due to compromised force sensor system alarm.The insulin pump was received with minor scratched lcd window, cracked case at the display window corners, cracked case at the reservoir tube window corners, cracked reservoir tube window, cracked belt clip slot, cracked battery tube threads, missing end cap sticker, stained address/serial number label and cracked reservoir tube lip.
|