MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAS
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/25/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The insulin pump passed the displacement test.No pump error noted.The insulin pump had compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or test for motor error alarm due to compromised force sensor system alarm.The insulin pump was received with minor scratched lcd window, cracked case at the display window corners, cracked case at the reservoir tube window corners, cracked reservoir tube window, cracked belt clip slot, cracked battery tube threads, missing end cap sticker, stained address/serial number label and cracked reservoir tube lip.
 
Event or Problem Description
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was unknown at the time of the incident.The customer also reported that the insulin was squirting during manual prime process.Customer was rewinding pump and during the prime it was spraying insulin.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7362829
Report Number3004209178-2018-59045
Device Sequence Number61388
Product Code OYC
UDI-Device Identifier00643169479630
UDI-Public(01)00643169479630
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA3523LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2017
Initial Report FDA Received Date03/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight140
-
-