• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4002C2860
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pusher assembly midjoint was retracted proximal to the introducer sheath friction lock.The embolization coil was intact with the pusher assembly.Conclusions: evaluation of the returned device revealed the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.The friction lock inner diameter (id) is smaller than the rest of the introducer sheath, which allows the introducer sheath to properly seat on the pusher assembly mid-joint.If the pusher assembly mid-joint is forcefully withdrawn beyond the introducer sheath friction lock, it is likely that resistance will be experienced by the physician upon attempting to advance the pc400.The non-penumbra microcatheter mentioned in the complaint was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a pelvic arteriovenous malformation (avm) using penumbra coils 400 (pc400¿s).It was noted that the patient¿s anatomy was moderately tortuous and slightly calcified.During the procedure, the physician placed five pc400¿s in the target vessel using a non-penumbra microcatheter.While attempting to advance a new pc400 into the microcatheter, the physician experienced resistance at the distal tip of the pc400¿s introducer sheath and the pc400 would not advance further; therefore, it was removed.The procedure was completed additional pc400¿s and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7363224
Report Number3005168196-2018-00611
Device Sequence Number13716117
Product Code HCG
UDI-Device Identifier00814548011981
UDI-Public00814548011981
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/08/2024
Device Catalogue Number4002C2860
Device Lot NumberF71048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/23/2018
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-