| Model Number |
METRIX AIR |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-18- user has high glucose value.Test strip (b)(4).Manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for hi blood glucose results.Grandson is calling on behalf of the customer.The customer is concerned with results obtained of hi, 336 and 202 mg/dl.The expected fasting blood glucose test result range is 120 mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a back to back blood test was performed non- fasting by the customer and produced test results of 508 and 493 mg/dl using true metrix air meter.The test strip lot manufacturer's expiration date is 4/24/2019 and open vial date is one month ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).
|
| |
|