• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125300-23
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.The reported material rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported material rupture appears to be related to circumstances of the procedure as it is likely the device interacted with the guiding catheter or anatomy during advancement resulting in the noted scratched balloon and reported material rupture.The treatment appears to be related to the operational context of the procedure as the stent was partially deployed and required extra dilatation to be fully deployed.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that following pre-dilatation, the xience alpine stent delivery system was advanced to the right coronary artery lesion without issue.Balloon inflation was attempted; however, the balloon ruptured at 6 atmospheres.The delivery system was removed without issue.As treatment, additional post-dilatation was performed on the implanted stent.The stent remains well apposed to the vessel wall at the intended site.There were no adverse patient effects and there was no clinically significant delay.There was no additional information provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7363417
Report Number2024168-2018-02083
Device Sequence Number1274310
Product Code NIQ
UDI-Device Identifier08717648199363
UDI-Public08717648199363
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2019
Device Catalogue Number1125300-23
Device Lot Number6082941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2018
Initial Report FDA Received Date03/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE WIRE: PROWATER
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient Weight69
-
-