The used device was returned to conmed for evaluation.Upon visual inspection, it was found that the device had a broken tip; the driver's fork was bent and broken.The likely cause of this failure is the application of excessive force to the device.This is a technique dependent device.A review of the manufacturing documents verified the device was produced per current and approved procedures and material specifications.Non-conformances regarding the product's identity, quality, safety, effectiveness or performance were not identified at the time of manufacture.A review of this lot number, with (b)(4) devices manufactured, showed no other complaint for this lot product and event description.(b)(4).The instructions for use advises the user of the following.Pull the driver straight out of bone.Do not rotate or oscillate driver about its axis during removal, or breakage of driver tip and/or anchor may result.Inspect instruments prior to use to ensure they are in good physical condition and function properly.There should be no loose, broken or misaligned parts.Inspect instruments after use to ensure they have not been damaged.Avoid unintended contact with other surgical instruments during use to prevent damage or breakage.Avoid lateral loading while inserting y-knot anchors.This issue will continue to be monitored through the complaint system to assure patient safety.
|
|
The conmed representative reported on behalf of the user facility that during a rc repair, one of the metal tips of the yrc03 driver broke off and the implant pulled out.The broken metal tip was retrieved with a grasper.The procedure was completed with an alternate yrc03 with a 7-minute surgical delay.No patient injury was reported.Upon gathering additional information, the reporter stated that no additional treatment or hospital stay was required due to this incident and the patient was recovering as expected.This report is raised on the basis of a reported malfunction with potential for injury with recurrence.
|