• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC SOLO HF CATHETER 5F RADIOLOGY BASIC TRAY (TRIPLE-LUMEN)(135CM); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC SOLO HF CATHETER 5F RADIOLOGY BASIC TRAY (TRIPLE-LUMEN)(135CM); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 3395335Q
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of reav2351 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (reav2351) have been reported from the same brazil facility.Device has not yet been returned for evaluation.
 
Event or Problem Description
It was reported that the catheter fractured.No other information was provided.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of fractured catheter was confirmed and the cause appears to be use related.The product returned for evaluation was one 5 fr dual lumen groshong nxt catheter.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through the use of syringes smaller than 10ml, by flushing against an occlusion, or due to excessive force applied during infusion procedures.The returned product sample was evaluated and a split was observed in the extension leg of the proximal (red) lumen, approximately 2 mm distal to the strain relief of the luer hub.This catheter damage was typical of a burst, and the characteristics observed which supported this type of failure included: 's' shaped split.Tensile weakness at the fracture site (due to material ballooning prior to burst).Granular fracture surface texture (typical of tearing failure modes).The nursing procedure manual can be found online at www.Bardaccess.Com and provides the following guidance: ¿do not use a syringe smaller than 10ml to flush or confirm patency.Patency should be assessed with a 10ml or larger syringe with sterile saline.Upon confirmation of patency, administration of medication should be given in a syringe appropriately sized for the does.Do not infuse against resistance.Infusion pressures should never exceed 25 psi.Smaller syringes generate more pressure than larger syringes.¿ the nursing procedure manual also provides guidance for re-establishing catheter patency in the event the catheter becomes occluded or resistance is felt during flushing.A lot history review (lhr) of reav2351 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (reav2351) have been reported from the same (b)(6) facility.
 
Event or Problem Description
It was reported that the catheter fractured.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO HF CATHETER 5F RADIOLOGY BASIC TRAY (TRIPLE-LUMEN)(135CM)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7364211
Report Number3006260740-2018-00505
Device Sequence Number1292083
Product Code LJS
UDI-Device Identifier00801741028373
UDI-Public(01)00801741028373
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2018
Device Model Number3395335Q
Device Catalogue Number3395335Q
Device Lot NumberREAV2351
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/19/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-