• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
Date of Event 12/06/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If explanted, give date: an explant was scheduled for (b)(6) 2018, however not yet confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis symfony multifocal intraocular lens (model zxr00, +24.0 diopter) was implanted in patient's operative eye on (b)(6) 2017.It was later scheduled to be explanted on (b)(6) 2018 in a secondary procedure because of patient experiencing blurry vision, halos, visual disturbance and light sensitivity.Reportedly the near vision was 20/25.No further information was provided.
 
Additional Manufacturer Narrative
During follow up, it was determined that the patient's date of birth is as follows: date of birth: (b)(6).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7364375
Report Number9614546-2018-00278
Device Sequence Number10806425
Product Code POE
UDI-Device Identifier05050474579170
UDI-Public(01)05050474579170(17)211110
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/10/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/10/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0240
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/21/2018
Supplement Date Received by Manufacturer03/23/2018
08/16/2018
Initial Report FDA Received Date03/22/2018
Supplement Report FDA Received Date04/22/2018
09/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Date Report Sent to FDA01/01/2005
-
-