| Model Number |
MMT-523NAP |
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
03/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the customer had low blood glucose level.The customer¿s blood glucose level was 53 mg/dl.The customer reported that the insulin pump received no delivery alarm and it was found during troubleshooting that the insulin exited.The customer reported that the no delivery alarm recurred.The customer reported that they tried all remedies.The customer was advised that the device would be replaced and agreed to return the insulin pump for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No unexpected no delivery or no reservoir alarms noted during testing.The device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.No unexpected prime or fill anomaly noted during testing.
|
| |
|