Catalog Number 8115539
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with 1) lumbar 5 spondylolysis.2) l3 -l4 vertebral endplate inflammation.3) lumbar 3, 4, 5 sacral vertebral disc herniation on and underwent pedicle screw fixation.Intra-op, when the doctor was tightening the screw of the cross-link, it broke.The doctor heard the broken sound when he used a screw driver to lock, and felt the screw was no longer rotating.Product came in contact with the patient.No fragment of the implant remained in the patient.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual and microscopic examinations were performed.The tip of one of the cross-link locking screws has fractured.A microscopic review of the fracture face gives an indication that there was a side load placed on the tip of the screw causing it to fracture as the screw was tightened.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7364874
Report Number1030489-2018-00442
Device Sequence Number19346116
Product Code KWP
UDI-Device Identifier00721902729690
UDI-Public00721902729690
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K030840
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8115539
Device Lot Number0555259W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Initial Date Received by Manufacturer 02/28/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date03/23/2018
Supplement Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
-
-