| Model Number |
ZXT225 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/05/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis symfony toric multifocal intraocular lens (iol) model zxt225, +24.5 diopter was explanted from patient's right eye in a secondary surgical procedure because the wrong toric power was implanted.Reportedly, the patient had irregular astigmatism pre-op, and a non-j&j system used to calculate the lens power intraoperatively, suggested the wrong lens power.The surgeon did not explant the lens right away because he was waiting to see if the patient was going to tolerate the lens.A zxt375 +27.5 diopter iol was then implanted as the replacement lens for the patient.No incision enlargement, no sutures were used, no vitrectomy was performed.The patient was reported doing fine when discharged.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 3/21/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|