MAUDE Adverse Event Report: ARTHREX INC. SCREW, CORTICAL; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX INC. SCREW, CORTICAL; PLATE, FIXATION, BONE Back to Search Results
Catalog Number AR-13380-44
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.An evaluation of the device cannot be performed as the device was not returned to arthrex.Additional information has been requested but not made available.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.The cause of the event is undetermined.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a distal femoral osteotomy procedure, as the surgeon was tightening the second cortical screw, ar-13380-44 (lot: 10130957, (b)(4)) into the contourlock femoral osteotomy plate, ar-13110l-02 (lot: 10162249, (b)(4)), the lateral cortex of the femur became fractured.The surgeon went on to insert the remaining screws (listed below) and then opened up a different manufacturer's device to hold the fracture and the plate together to provide support so that the fracture did not spread.A fluoro was taken during the case to confirm the fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW, CORTICAL
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7365039
Report Number1220246-2018-00099
Device Sequence Number1949603
Product Code HRS
UDI-Device Identifier00888867011571
UDI-Public00888867011571
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K032187
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/01/2022
Device Catalogue NumberAR-13380-44
Device Lot Number10130957
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2018
Initial Report FDA Received Date03/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AR-13110L-02, FEMORAL OSTEO PLATE, LOT 10162249.
Outcome Attributed to Adverse Event Other;
-
-