| Model Number |
N/A |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Foreign source - (b)(6).Customer has indicated that the product is in process of being returned to orthopediatrics for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It has been reported that following the close of a derotational osteotomy procedure, the infant locking cannulated blade plate was found, via x-ray, to be bent.The patient was re-opened and the infant blade plate replaced with a child blade plate.No additional patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Photographs of x-rays confirmed that the plate was bent.The device was not returned for evaluation.The device was made to specification and there were no manufacturing deviations.There have been no recent design changes.A root cause was unable to be determined.Per the system instructions for use, a risk associated with the use of this device is "bending, fracture, or migration of the implant".If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Orthopediatrics will continue to monitor for trends.
|
| |
|