• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381134
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that before use a bd angiocath¿ iv catheter was found with foreign matter as ¿the catheter has a fringe close to the tip.¿ there was no report of exposure, injury or medical intervention needed.
 
Additional Manufacturer Narrative
Investigation results: samples or photos have not received for analysis of the incident in question.Our quality engineer was unable to verify the reported complaint.Dhr review: the assembly lot: 7116857, used in the claimed final product lot: 7116857 produced from 04/29/2017 to 05/31/2017 in acam #4, used in the claimed final product lot: 7117509 of angiocath 20ga x 1.16 was checked as testes of presence as "foreign / no foreign matter" and no records of this defect were found.Qn/ ncmr review: there are no records of quality notification (qn) or non-conformance report (rnc) of foreign matter, for the lot involved in this complaint, according to the history of the lot above.Because no objective evidence of this issue was found during the analysis of the device history record and quality notifications for the claimed lot, it was not possible to confirm this complaint as being generated by manufacturing process.Based on the investigation results to date the root cause was not determined.Complaints received for this device and the reported defect will continue to be tracked and trended.The information will be captured in the trend reports and monitored monthly.Collected data are regularly reviewed to identify emerging trends.Based on this, a capa is not needed at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ANGIOCATH¿ IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
MDR Report Key7366934
Report Number9610048-2018-00030
Device Sequence Number1144146
Product Code FOZ
UDI-Device Identifier30382903811343
UDI-Public30382903811343
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2022
Device Catalogue Number381134
Device Lot Number7117509
Initial Date Received by Manufacturer 03/15/2018
Supplement Date Received by Manufacturer03/15/2018
Initial Report FDA Received Date03/23/2018
Supplement Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-