• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS DOT PROV 4F DL FULL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS DOT PROV 4F DL FULL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Catalog Number S1274108
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of rebu1047 showed three other similar product complaint(s) from this lot number.The complaints for this lot number (rebu1047) have been reported from one facility.
 
Event or Problem Description
It was reported that when the vas team arrived to assess the patient¿s picc, the patient's dressing had been removed and the lumens were wrapped in an abdominal pad due to leaking.It was noted that it was fractured ¿where the tubing enters into the end portion of the lumens¿ and that the fracture was visible.It was also noted that ¿blood was leaking out.¿ the picc was removed and a sterile dressing was placed.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is confirmed.An initial visual observation showed use residue throughout the sample and a needleless injection cap was observed to be attached to each luer hub and an end cap was attached to the needless injection cap on the purple lumen.A split was observed in the extension tubing of the purple lumen just within the distal end of the luer connector.The ink on the extension legs was observed to be slightly worn and faded and the pinch clamp of the red lumen was observed to be engaged.A microscopic observation revealed the fracture surface of the split in the extension tubing of the purple lumen was rough and exhibited cracks/fissures and beach marks.
 
Event or Problem Description
It was reported that when the vas team arrived to assess the patient¿s picc, the patient's dressing had been removed and the lumens were wrapped in an abdominal pad due to leaking.It was noted that it was fractured ¿where the tubing enters into the end portion of the lumens¿ and that the fracture was visible.It was also noted that ¿blood was leaking out.¿ the picc was removed and a sterile dressing was placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV 4F DL FULL 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7367107
Report Number3006260740-2018-00515
Device Sequence Number2312409
Product Code LJS
UDI-Device Identifier00801741130120
UDI-Public(01)00801741130120
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other,Notification
Type of Report Initial,Followup
Report Date (Section B) 02/12/2019
Report Date (Section F) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberS1274108
Device Lot NumberREBU1047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Event Location Hospital
Date Report to Manufacturer02/26/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/05/2019
Initial Report FDA Received Date03/23/2018
Supplement Report FDA Received Date02/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-