Catalog Number B2120-040
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: a visual inspection was performed on the returned device.The reported balloon rupture and balloon tear were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot and relabeled lots that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot or relabeled lots.The investigation was unable to determine a conclusive cause for the reported balloon rupture.The balloon tear appears to be related to operational context of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure of the heavily tortuous, heavily calcified, eccentric, de novo, mid common iliac, the armada 35 balloon was inflated at 10 atmospheres (atm), but ruptured.The device was removed from the anatomy without reported issue.There was no reported adverse patient effect or a clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7367529
Report Number2024168-2018-02138
Device Sequence Number13726862
Product Code LIT
UDI-Device Identifier08717648155550
UDI-Public08717648155550
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Catalogue NumberB2120-040
Device Lot Number40904G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Initial Report FDA Received Date03/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-