Model Number S7 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.A medtronic representative went to site to test the equipment.Representative reported that cmos battery was replaced and over 300 patients were removed from the system.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect part has not been received by the manufacturer for evaluation.
|
|
Event Description
|
A site representative reported that during cranial biopsy, the navigation system unexpectedly became unresponsive while in pre-planning and registration screens.System was hard restarted and issue resolved.The procedure was completed with the use of navigation.There was no delay to procedure.No known impact on patient outcome.
|
|
Manufacturer Narrative
|
Correction: device code updated with correct information.If information is provided in the future, a supplemental report will be issued.
|
|