• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551PNAB
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 10/09/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Unit received with completely broken reservoir tube lip.Unable to perform basic occlusion test, displacement tes and occlusion test due to broken reservoir tube lip.However, unit was received passed rewind test, prime/a33 test and excessive no delivery test.Unit had minor scratched lcd window, cracked lcd window, cracked battery tube threads, missing reservoir tube o-ring, cracked case at display window corners, cracked reservoir tube, cracked reservoir tube window, cracked case at reservoir tube window corners, stained address/serial number label and stained end cap sticker.
 
Event or Problem Description
The customer reported via phone call that pump had crack on the side of the reservoir and experienced high blood glucose level.The customer¿s blood glucose level was 307 mg/dl.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551PNAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7368318
Report Number3004209178-2018-59617
Device Sequence Number16282416
Product Code OZO
UDI-Device Identifier00643169507265
UDI-Public(01)00643169507265
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551PNAB
Device Catalogue NumberMMT-551PNAB
Device Lot NumberA3551PNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Initial Report FDA Received Date03/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight140
-
-