• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 26 MM +4; PRSTHSIS,HIP,SMI-CONSTRAIND,MTAL/CRAMIC/POLYMR,CMNTDORNN-PROUS,UNCMNTD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 26 MM +4; PRSTHSIS,HIP,SMI-CONSTRAIND,MTAL/CRAMIC/POLYMR,CMNTDORNN-PROUS,UNCMNTD Back to Search Results
Catalog Number 71342604
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/22/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to infection.The tandem bipolar shell/ xlpe liner and the femoral head were removed.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and although no clinically relevant supporting documentation was provided to perform a thorough medical investigation, it was reported that the revision was required approximately 3 weeks post implantation due to infection.It was further communicated that the surgeon did not fault the device.The patient impact beyond the likely administration of antibiotic therapy and the revision procedure itself cannot be determined.No further medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 26 MM +4
Common Device Name
PRSTHSIS,HIP,SMI-CONSTRAIND,MTAL/CRAMIC/POLYMR,CMNTDORNN-PROUS,UNCMNTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7368748
Report Number1020279-2018-00482
Device Sequence Number1705226
Product Code LZO
UDI-Device Identifier03596010488909
UDI-Public03596010488909
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71342604
Device Lot Number17JM17369
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/07/2018
Supplement Date Received by Manufacturer03/07/2018
Initial Report FDA Received Date03/25/2018
Supplement Report FDA Received Date06/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-