| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.Husband is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 281 and 205 mg/dl.The customer's expected fasting blood glucose test result range is 125 - 130 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 127 mg/dl and 131 mg/dl using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 02/21/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :191mg/dl date: (b)(6) 2018 time:03:20pm fasting; result 2 :281mg/dl date: (b)(6) 2018 time:03:17pm fasting; result 3 :155mg/dl date: (b)(6) 2018 time:03:15pm fasting; result 4 :131mg/dl date: (b)(6) 2018 time:09:35am fasting; result 5 :205mg/dl date: (b)(6) 2018 time:07:38am fasting.
|
| |
|
Search Alerts/Recalls
|