U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.Husband is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 281 and 205 mg/dl.The customer's expected fasting blood glucose test result range is 125 - 130 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 127 mg/dl and 131 mg/dl using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 02/21/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :191mg/dl date: (b)(6) 2018 time:03:20pm fasting; result 2 :281mg/dl date: (b)(6) 2018 time:03:17pm fasting; result 3 :155mg/dl date: (b)(6) 2018 time:03:15pm fasting; result 4 :131mg/dl date: (b)(6) 2018 time:09:35am fasting; result 5 :205mg/dl date: (b)(6) 2018 time:07:38am fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7369599
Report Number1000113657-2018-00352
Device Sequence Number37523
Product Code NBW
UDI-Device Identifier00021292007065
UDI-Public(01)00021292007065
Combination Product (Y/N)N
Initial Reporter StateMD
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/21/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMU2593
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2018
Initial Report FDA Received Date03/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-