| Model Number |
N/A |
| Medical Device Problem Code |
Kinked (1339)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Blood Loss (2597)
|
| Date of Event |
03/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was not implanted; device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code has been referenced in the conclusions section.Conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the 5 fr.Pinnacle sheath kinked in an artery during a left heart cath.It was a single stick.The entry angle was steeper than 45%, which probably contributed to the kink.The physician had difficulty advancing and torqueing the catheters.The sheath was switched out for an unkinked sheath and the procedure was completed successfully with no injury to patient.Additional information was received on 3/7/18: it was reported that there was a minimal blood loss when getting access.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The evaluation of the actual device could not be conducted due to the device not being returned.
|
| |
|
Search Alerts/Recalls
|