The original sample was not returned.Pictures of the fragment removed from the patient were provided.The pictures confirmed the piece to be a portion of the simpulse device.There was a piece of the irrigation connector tube securely bonded into the showerhead tip body.The irrigation connector tube had fractured just below the bonded joint.Based on the pictures provided, it appears the component may have fractured due to the forces applied to the device by the user during use.There is no way for us to conclusively identify what condition or combination of conditions were created to produce that force.A lot number was not provided, therefore, a review of the manufacturing records could not be performed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Not returned for evaluation.
|