Model Number 728306 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
We have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
|
|
Event Description
|
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported it was difficult to hear the patient.It was confirmed with the philips field service engineer (fse) the customer could not hear the patient from the gantry microphone.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.
|
|
Manufacturer Narrative
|
The customer reported on (b)(6) 2018 that the operator could not hear sound coming from the microphone gantry at all times.However, the customer confirmed that the operator was able to hear the patients thru the open door.Additionally, the customer confirmed that the patient could hear the operator give manual instructions from the ct box and could also hear the auto-voice commands.The customer confirmed with the philips helpdesk (when this issue was reported) that there was no patient involvement and no harm as a result of this event.The philips field service engineer (fse) went to the customer site on 08-mar-2018 to evaluate the ct system.The fse confirmed the customer¿s allegation and found that the microphone had failed.The fse replaced two breathing light assemblies to correct and resolve the issue.The system was handed back to the customer for clinical use.The system is operational and in clinical use.This event has been determined not to be a reportable event.
|
|
Search Alerts/Recalls
|
|
|