• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Device Problem No Audible Alarm (1019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
We have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported it was difficult to hear the patient.It was confirmed with the philips field service engineer (fse) the customer could not hear the patient from the gantry microphone.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported on (b)(6) 2018 that the operator could not hear sound coming from the microphone gantry at all times.However, the customer confirmed that the operator was able to hear the patients thru the open door.Additionally, the customer confirmed that the patient could hear the operator give manual instructions from the ct box and could also hear the auto-voice commands.The customer confirmed with the philips helpdesk (when this issue was reported) that there was no patient involvement and no harm as a result of this event.The philips field service engineer (fse) went to the customer site on 08-mar-2018 to evaluate the ct system.The fse confirmed the customer¿s allegation and found that the microphone had failed.The fse replaced two breathing light assemblies to correct and resolve the issue.The system was handed back to the customer for clinical use.The system is operational and in clinical use.This event has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7371213
MDR Text Key103662126
Report Number1525965-2018-00378
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/01/2018
Initial Date FDA Received03/26/2018
Supplement Dates Manufacturer Received03/01/2018
Supplement Dates FDA Received08/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-