| Model Number |
1430US |
| Medical Device Problem Codes |
Device Inoperable (1663); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Concomitant medical products: 1103 vad implanted: (b)(6) 2017, ddbc3d1 icd, implanted: (b)(6) 2013.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the controller ac adapter did not provide power.The ac adapter was exchanged.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the controller ac adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis revealed that the device passed functional testing and visual inspection.The ac adapter was able to provide power as intended.As a result, the reported event was not confirmed.Based on the risk documentation, possible root causes of the reported event may be attributed to multiple factors, including but not limited to a marginal connection of the controller ac adapter to the wall outlet and/or a marginal connection of the controller ac adapter to the controller.If information is provided in the future, a supplemental report will be issued.
|
| |
|