MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ROTICULATOR; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ROTICULATOR; STAPLE, IMPLANTABLE Back to Search Results
Model Number 017614
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/07/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: occurred during a laparoscopic low anterior resection (lar) procedure.The stapling devices were being used for the rectum resection.The patient had undergone radiation treatments so the tissue was very thick.The patient was also obese with a very narrow pelvis.The powered stapling handle broke due to the thickness of the rectum.The surgeon forced the trigger pushing intensively.There was also a problem unloading the powered stapling device.Seems as if the axis is outside the adapter due to the extreme force the surgeon did try to place and fire the stapler with the reload.A component of the inner adapter axis broke off.Then tried to performed with resection using a manual stapling handle with another reload, but it did not work either.Then tried with a roticulator stapler, but the staple line was not continuous.Was not safe to use a circular stapler in order to perform the anastomosis.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTICULATOR
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7371847
Report Number2647580-2018-01520
Device Sequence Number39655
Product Code GDW
UDI-Device Identifier10884521047709
UDI-Public10884521047709
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K855047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number017614
Device Catalogue Number017614
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2018
Initial Report FDA Received Date03/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-