MASIMO - 40 PARKER SEDLINE EEG SENSOR; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 4248-9
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported "high psi value was displayed by root while patient was enough anesthetized." no patient impact or consequences were reported.
 
Additional Manufacturer Narrative
The returned sensor was evaluated.During evaluation the sensor passed external visual inspection.The sensor failed continuity testing due to an open r1, r2, l2, ct, and cb across the electrodes.Functional testing could not be performed as the sensor was returned used and could not be tested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEDLINE EEG SENSOR
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key7372419
Report Number2031172-2018-00168
Device Sequence Number19090437
Product Code MWI
UDI-Device Identifier10843997005804
UDI-Public10843997005804
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K151644
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2019
Device Model Number4248-9
Device Catalogue Number2479
Device Lot Number17N55
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 03/02/2018
Supplement Date Received by Manufacturer04/17/2018
Initial Report FDA Received Date03/26/2018
Supplement Report FDA Received Date05/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-