SMITH & NEPHEW, INC. ANTHOLOGY ANTERIOR BROACH SZ11; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
71365211 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported a 2 hour delay in surgery as consequence of a broken broach handle.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were returned and evaluated.A visual inspection of the returned broaches and trial necks indicated that they all showed signs of wear from use.One of the broaches was fractured at the connector base.The fractured off portion was returned.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A clinical analysis indicated that no further clinically relevant information was provided, therefore the prior clinical/medical comment will remain unchanged as the product findings support the previously submitted comment.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint device was not returned.The photo provided confirmed the broach is broken at the connector.The clinical/medical team concluded, per complaint and correspondence, although the broaches are all ¿in the same shape¿ the procedure was completed with a backup device without report of patient injury after the 2 hour surgical delay.The ¿shape¿ (condition) of the broach was likely a contributing factor to the reported breakage and subsequent surgical delay.The ifc, maintenance, cleaning and sterilization of s&n orthopaedic devices brochure_01004_v2 indicates that hospital trained personnel should visually inspect all reusable devices (for damage, wear, maintenance) and to discard blunt or damaged instruments which is in alignment with common operating room/standard operating procedures.The patient impact beyond the reported 2 hour surgical delay cannot be determined as the current patient status remains unknown; however, no patient injury was reported.No further medical assessment can be rendered at this time.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|