Model Number ZCT225
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was inserted and removed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zct225 22.5 diopter intraocular lens (iol) was inserted and removed from the patient's left eye (os) because the lens was noticed fractured once implanted.There was no incision enlargement, no vitrectomy, and no sutures used.The replacement lens was a back up lens of the same model and diopter.Reportedly, there was no patient injury and the patient is doing fine post-operatively.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 03/20/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection showed the lens was cut in the middle, most probably to make remove and replacement possible, apart from the detached haptic.Due to the condition of the lens, further investigation could not be performed on the sample.Complaint event of lens damaged is confirmed, however the cause is unknown.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency.Complaint event of lens damaged is confirmed, however the cause is unknown.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7373158
Report Number3011852734-2018-00033
Device Sequence Number1144135
Product Code HQL
UDI-Device Identifier05050474552784
UDI-Public(01)05050474552784(17)210722
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/22/2021
Device Model NumberZCT225
Device Catalogue NumberZCT225U225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Initial Date Received by Manufacturer 02/26/2018
Supplement Date Received by Manufacturer04/09/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-