Model Number VLOCM0625
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic procedure, while being applied for the suturing of vaginal blunt, the device¿s thread broke.Surgeon tried to use other two sutures but had the same issue.Another device was used in order to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of nine devices.The visual inspection and functional evaluation of the sample had acceptable results.A review of the device history record indicates the product was released meeting all manufacturers¿ quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 90
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7373161
Report Number1219930-2018-01689
Device Sequence Number11038706
Product Code GAM
UDI-Device Identifier20884521132631
UDI-Public20884521132631
Combination Product (Y/N)N
Initial Reporter CountryRO
PMA/510(K) Number
K100257
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Model NumberVLOCM0625
Device Catalogue NumberVLOCM0625
Device Lot NumberA5G1353VWX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2018
Initial Date Received by Manufacturer 03/09/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date06/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-