• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint (b)(4).
 
Event or Problem Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported that the intercom system failed.The philips field service engineer (fse) confirmed that the operator could not hear the patient at all times.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.
 
Additional Manufacturer Narrative
The customer reported on (b)(6) 2018 that the intercom system failed and as a result, the doctor could not hear the patient voices from ct box.The customer confirmed with the philips helpdesk (when this issue was reported) that there was no patient impact and no harm as a result of this event.The philips field service engineer (fse) arrived at the customer site on (b)(6) 2018 to evaluate the ct system.The fse tested the breathing light board (assembly) and determined that it was not functioning.The fse replaced the breathing light assembly to correct and resolve the issue.The system was tested successfully and handed back to the customer for clinical use.The system is operational and in clinical use.This event has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7373347
Report Number1525965-2018-00380
Device Sequence Number106964
Product Code JAK
UDI-Device Identifier00884838059474
UDI-Public(01)00884838059474
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K060937
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/03/2018
Supplement Date Received by Manufacturer03/03/2018
Initial Report FDA Received Date03/27/2018
Supplement Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-